Cyanide Cell Test
Catalogue Number | 114561 |
Overview | The method corresponds to DIN 38405-13 and is analogous to ISO 6703, EPA 335.2, APHA 4500-CN-E, and ASTM D2036-09 D. USEPA approved for wastewater. This Spectroquant® Cyanide Cell Test allows the accurate quantification of the free and readily liberated cyanide content in aqueous samples. Method applied: Cyanide ions react with a chlorinating agent to form cyanogen chloride which together with 1,3-dimethylbarbituric acid forms a violet dye (pyridine-free König reaction) that is determined photometrically. The Spectroquant® Cell Tests come with prefilled 16 mm round cells and all the required reagents to perform the analysis according to the instruction leaflet provided. All Spectroquant® Cell and Reagent Tests are equipped with the unique Live ID which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates. |
Product Information | |
---|---|
HS Code | 3822 19 00 |
Quality Level | MQ100 |
Storage and Shipping Information | |
---|---|
Storage | Store at +15°C to +25°C. |